MIC-004
2-3" outside micrometer
- Serial
- 71104455
- Manufacturer
- Mitutoyo
- Location
- Cell 2
- Owner
- Luis Ortega
- Interval
- 1 yr
- Next due
- Oct 18, 2027
- Status
- Active
| Cal date | Source | By | Master / lab | Cert no. | File | As found | As left | Due | Entry |
|---|---|---|---|---|---|---|---|---|---|
| Oct 18, 2026 | In-house | LO | GBS-01 | Out of tolerance | Adjusted | Oct 18, 2027 | #76 | ||
| Oct 18, 2025 | In-house | LO | GBS-01 | In tolerance | In tolerance | Oct 18, 2026 | #31 |
Out-of-tolerance reviews
Opened Oct 18, 2026, open. Jobs 24-1187 and 24-1192 measured with MIC-004 since its last good calibration. Pulled 6 parts from 24-1192 for re-inspection on the CMM; results pending.
Entries are never edited; a correction is a new entry that names the one it corrects. Generated by Gauge Recall.